TY - JOUR A1 - Gräßl, Andreas A1 - Renz, Wolfgang A1 - Hezel, Fabian A1 - Dieringer, Matthias A. A1 - Winter, Lukas A1 - Özerdem, Celal A1 - Rieger, Jan A1 - Kellmann, Peter A1 - Santoro, Davide A1 - Lindel, Tomasz D. A1 - Frauenrath, Tobias A1 - Pfeiffer, Harald A1 - Niendorf, Thoralf T1 - Modular 32-channel transceiver coil array for cardiac MRI at 7.0T JF - Magnetic Resonance in Medicine N2 - Purpose To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T. Methods The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study. Results Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly. Conclusion The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T. Y1 - 2013 U6 - https://doi.org/10.1002/mrm.24903 SN - 1522-2594 VL - 72 IS - 1 SP - 276 EP - 290 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Frauenrath, Tobias A1 - Fuchs, Katharina A1 - Dieringer, Matthias A. A1 - Özerdem, Celal A1 - Patel, Nishan A1 - Renz, Wolfgang A1 - Greiser, Andreas A1 - Elgeti, Thomas A1 - Niendorf, Thoralf T1 - Detailing the use of magnetohydrodynamic effects for synchronization of MRI with the cardiac cycle: A feasibility study JF - Journal of Magnetic Resonance Imaging N2 - Purpose: To investigate the feasibility of using magnetohydrodynamic (MHD) effects for synchronization of magnetic resonance imaging (MRI) with the cardiac cycle. Materials and Methods: The MHD effect was scrutinized using a pulsatile flow phantom at B0 = 7.0 T. MHD effects were examined in vivo in healthy volunteers (n = 10) for B0 ranging from 0.05–7.0 T. Noncontrast-enhanced MR angiography (MRA) of the carotids was performed using a gated steady-state free-precession (SSFP) imaging technique in conjunction with electrocardiogram (ECG) and MHD synchronization. Results: The MHD potential correlates with flow velocities derived from phase contrast MRI. MHD voltages depend on the orientation between B0 and the flow of a conductive fluid. An increase in the interelectrode spacing along the flow increases the MHD potential. In vivo measurement of the MHD effect provides peak voltages of 1.5 mV for surface areas close to the common carotid artery at B0 = 7.0 T. Synchronization of MRI with the cardiac cycle using MHD triggering is feasible. MHD triggered MRA of the carotids at 3.0 T showed an overall image quality and richness of anatomic detail, which is comparable to ECG-triggered MRAs. Conclusion: This feasibility study demonstrates the use of MHD effects for synchronization of MR acquisitions with the cardiac cycle. J. Magn. Reson. Imaging 2012;36:364–372. © 2012 Wiley Periodicals, Inc. Y1 - 2012 U6 - https://doi.org/10.1002/jmri.23634 SN - 1522-2586 VL - 36 IS - 2 SP - 364 EP - 372 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Grande, Marion A1 - Meffert, Elisabeth A1 - Schoenberger, Eva A1 - Jung, Stefanie A1 - Frauenrath, Tobias A1 - Huber, Walter A1 - Hussmann, Katja A1 - Moormann, Mareike A1 - Heim, Stefan T1 - From a concept to a word in a syntactically complete sentence: An fMRI study on spontaneous language production in an overt picture description task JF - NeuroImage N2 - Spontaneous language has rarely been subjected to neuroimaging studies. This study therefore introduces a newly developed method for the analysis of linguistic phenomena observed in continuous language production during fMRI. Most neuroimaging studies investigating language have so far focussed on single word or — to a smaller extent — sentence processing, mostly due to methodological considerations. Natural language production, however, is far more than the mere combination of words to larger units. Therefore, the present study aimed at relating brain activation to linguistic phenomena like word-finding difficulties or syntactic completeness in a continuous language fMRI paradigm. A picture description task with special constraints was used to provoke hesitation phenomena and speech errors. The transcribed speech sample was segmented into events of one second and each event was assigned to one category of a complex schema especially developed for this purpose. The main results were: conceptual planning engages bilateral activation of the precuneus. Successful lexical retrieval is accompanied – particularly in comparison to unsolved word-finding difficulties – by the left middle and superior temporal gyrus. Syntactic completeness is reflected in activation of the left inferior frontal gyrus (IFG) (area 44). In sum, the method has proven to be useful for investigating the neural correlates of lexical and syntactic phenomena in an overt picture description task. This opens up new prospects for the analysis of spontaneous language production during fMRI. Y1 - 2012 U6 - https://doi.org/10.1016/j.neuroimage.2012.03.087 SN - 1522-2586 VL - 61 IS - 3 SP - 702 EP - 714 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Martin, Conrad A1 - Frauenrath, Tobias A1 - Özerdem, Celal A1 - Renz, Wolfgang A1 - Niendorf, Thoralf T1 - Development and evaluation of a small and mobile Magneto Alert Sensor (MALSE) to support safety requirements for magnetic resonance imaging JF - European Radiology N2 - Objective The purpose of this study is to (i) design a small and mobile Magnetic field ALert SEnsor (MALSE), (ii) to carefully evaluate its sensors to their consistency of activation/deactivation and sensitivity to magnetic fields, and (iii) to demonstrate the applicability of MALSE in 1.5 T, 3.0 T and 7.0 T MR fringe field environments. Methods MALSE comprises a set of reed sensors, which activate in response to their exposure to a magnetic field. The activation/deactivation of reed sensors was examined by moving them in/out of the fringe field generated by 7TMR. Results The consistency with which individual reed sensors would activate at the same field strength was found to be 100% for the setup used. All of the reed switches investigated required a substantial drop in ambient magnetic field strength before they deactivated. Conclusions MALSE is a simple concept for alerting MRI staff to a ferromagnetic object being brought into fringe magnetic fields which exceeds MALSEs activation magnetic field. MALSE can easily be attached to ferromagnetic objects within the vicinity of a scanner, thus creating a barrier for hazardous situations induced by ferromagnetic parts which should not enter the vicinity of an MR-system to occur. KW - MRI KW - MR safety KW - Magneto alert sensor KW - High field MRI KW - Uktrahigh field MRI Y1 - 2011 U6 - https://doi.org/10.1007/s00330-011-2153-z SN - 1432-1084 VL - 21 SP - 2187 EP - 2192 PB - Springer CY - Berlin, Heidelberg ER - TY - JOUR A1 - Dieringer, Matthias A. A1 - Renz, Wolfgang A1 - Lindel, Tomasz D. A1 - Seifert, Frank A1 - Frauenrath, Tobias A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Waiczies, Helmar A1 - Hoffmann, Werner A1 - Rieger, Jan A1 - Pfeiffer, Harald A1 - Ittermann, Bernd A1 - Schulz-Menger, Jeanette A1 - Niendorf, Thoralf T1 - Design and application of a four-channel transmit/receive surface coil for functional cardiac imaging at 7T JF - Journal of Magnetic Resonance Imaging N2 - Purpose To design and evaluate a four-channel cardiac transceiver coil array for functional cardiac imaging at 7T. Materials and Methods A four-element cardiac transceiver surface coil array was developed with two rectangular loops mounted on an anterior former and two rectangular loops on a posterior former. specific absorption rate (SAR) simulations were performed and a Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-1 calibration method was applied prior to obtain 2D FLASH CINE (mSENSE, R = 2) images from nine healthy volunteers with a spatial resolution of up to 1 × 1 × 2.5 mm3. Results Tuning and matching was found to be better than 10 dB for all subjects. The decoupling (S21) was measured to be >18 dB between neighboring loops, >20 dB for opposite loops, and >30 dB for other loop combinations. SAR values were well within the limits provided by the IEC. Imaging provided clinically acceptable signal homogeneity with an excellent blood-myocardium contrast applying the Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-2 calibration approach. Conclusion A four-channel cardiac transceiver coil array for 7T was built, allowing for cardiac imaging with clinically acceptable signal homogeneity and an excellent blood-myocardium contrast. Minor anatomic structures, such as pericardium, mitral, and tricuspid valves and their apparatus, as well as trabeculae, were accurately delineated. Y1 - 2011 U6 - https://doi.org/10.1002/jmri.22451 SN - 1522-2586 VL - 33 IS - 3 SP - 736 EP - 741 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Schopp, Christoph A1 - Rohrbach, Felix A1 - Langer, Luc A1 - Heuermann, Holger T1 - Detection of welding wire length by active S11 measurement JF - IEEE Transactions on Plasma Science N2 - A novel method to determine the extruded length of a metallic wire for a directed energy deposition (DED) process using a microwave (MW) plasma jet with a straight-through wire feed is presented. The method is based on the relative comparison of the measured frequency response obtained by the large-signal scattering parameter (Hot-S) technique. In the practical working range, repeatability of less than 6% for a nonactive plasma and 9% for the active plasma state is found. Measurements are conducted with a focus on a simple solution to decrease the processing time and reduce the integration time of the process into the existing hardware. It is shown that monitoring a single frequency for magnitude and phase changes is sufficient to achieve good accuracy. A combination of different measurement values to determine the length is possible. The applicability to different diameter of the same material is shown as well as a contact detection of the wire and metallic substrate. KW - Circuit simulation KW - Hot S-parameter KW - Modeling KW - Plasma KW - Plasma diagnostics Y1 - 2024 U6 - https://doi.org/10.1109/TPS.2024.3356659 SN - 0093-3813 (Print) SN - 1939-9375 (Online) IS - Early Access SP - 1 EP - 6 PB - IEEE ER - TY - JOUR A1 - Turdumamatov, Samat A1 - Belda, Aljoscha A1 - Heuermann, Holger T1 - Shaping a decoupled atmospheric pressure microwave plasma with antenna structures, Maxwell’s equations, and boundary conditions JF - IEEE Transactions on Plasma Science N2 - This article addresses the need for an innovative technique in plasma shaping, utilizing antenna structures, Maxwell’s laws, and boundary conditions within a shielded environment. The motivation lies in exploring a novel approach to efficiently generate high-energy density plasma with potential applications across various fields. Implemented in an E01 circular cavity resonator, the proposed method involves the use of an impedance and field matching device with a coaxial connector and a specially optimized monopole antenna. This setup feeds a low-loss cavity resonator, resulting in a high-energy density air plasma with a surface temperature exceeding 3500 o C, achieved with a minimal power input of 80 W. The argon plasma, resembling the shape of a simple monopole antenna with modeled complex dielectric values, offers a more energy-efficient alternative compared to traditional, power-intensive plasma shaping methods. Simulations using a commercial electromagnetic (EM) solver validate the design’s effectiveness, while experimental validation underscores the method’s feasibility and practical implementation. Analyzing various parameters in an argon atmosphere, including hot S -parameters and plasma beam images, the results demonstrate the successful application of this technique, suggesting its potential in coating, furnace technology, fusion, and spectroscopy applications. KW - 3-D printing KW - Furnace KW - Fusion KW - Hot S-parameter KW - Mode converter Y1 - 2024 U6 - https://doi.org/10.1109/TPS.2024.3383589 SN - 0093-3813 (Print) SN - 1939-9375 (Online) IS - Early Access SP - 1 EP - 9 PB - IEEE ER - TY - JOUR A1 - Fiedler, Thomas M. A1 - Orzada, Stephan A1 - Flöser, Martina A1 - Rietsch, Stefan H. G. A1 - Quick, Harald H. A1 - Ladd, Mark E. A1 - Bitz, Andreas T1 - Performance analysis of integrated RF microstrip transmit antenna arrays with high channel count for body imaging at 7 T JF - NMR in Biomedicine N2 - The aim of the current study was to investigate the performance of integrated RF transmit arrays with high channel count consisting of meander microstrip antennas for body imaging at 7 T and to optimize the position and number of transmit ele- ments. RF simulations using multiring antenna arrays placed behind the bore liner were performed for realistic exposure conditions for body imaging. Simulations were performed for arrays with as few as eight elements and for arrays with high channel counts of up to 48 elements. The B1+ field was evaluated regarding the degrees of freedom for RF shimming in the abdomen. Worst-case specific absorption rate (SARwc ), SAR overestimation in the matrix compression, the number of virtual obser- vation points (VOPs) and SAR efficiency were evaluated. Constrained RF shimming was performed in differently oriented regions of interest in the body, and the devia- tion from a target B1+ field was evaluated. Results show that integrated multiring arrays are able to generate homogeneous B1+ field distributions for large FOVs, espe- cially for coronal/sagittal slices, and thus enable body imaging at 7 T with a clinical workflow; however, a low duty cycle or a high SAR is required to achieve homoge- neous B1+ distributions and to exploit the full potential. In conclusion, integrated arrays allow for high element counts that have high degrees of freedom for the pulse optimization but also produce high SARwc , which reduces the SAR accuracy in the VOP compression for low-SAR protocols, leading to a potential reduction in array performance. Smaller SAR overestimations can increase SAR accuracy, but lead to a high number of VOPs, which increases the computational cost for VOP evaluation and makes online SAR monitoring or pulse optimization challenging. Arrays with interleaved rings showed the best results in the study. KW - body imaging at UHF MRI KW - integrated transmit coil arrays KW - VOP compression Y1 - 2021 U6 - https://doi.org/10.1002/nbm.4515 SN - 0952-3480 (ISSN) SN - 1099-1492 (eISSN) VL - 34 IS - 7 PB - Wiley CY - Weinheim ER - TY - JOUR A1 - Frauenrath, Tobias A1 - Hezel, Fabian A1 - Renz, Wolfgang A1 - de Geyer d'Orth, Thibaut A1 - Dieringer, Matthias A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Prothmann, Marcel A1 - Schulz-Menger, Jeanette A1 - Niendorf, Thoralf T1 - Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla JF - Journal of Cardiovascular Magnetic Resonance N2 - Background To demonstrate the applicability of acoustic cardiac triggering (ACT) for imaging of the heart at ultrahigh magnetic fields (7.0 T) by comparing phonocardiogram, conventional vector electrocardiogram (ECG) and traditional pulse oximetry (POX) triggered 2D CINE acquisitions together with (i) a qualitative image quality analysis, (ii) an assessment of the left ventricular function parameter and (iii) an examination of trigger reliability and trigger detection variance derived from the signal waveforms. Results ECG was susceptible to severe distortions at 7.0 T. POX and ACT provided waveforms free of interferences from electromagnetic fields or from magneto-hydrodynamic effects. Frequent R-wave mis-registration occurred in ECG-triggered acquisitions with a failure rate of up to 30% resulting in cardiac motion induced artifacts. ACT and POX triggering produced images free of cardiac motion artefacts. ECG showed a severe jitter in the R-wave detection. POX also showed a trigger jitter of approximately Δt = 72 ms which is equivalent to two cardiac phases. ACT showed a jitter of approximately Δt = 5 ms only. ECG waveforms revealed a standard deviation for the cardiac trigger offset larger than that observed for ACT or POX waveforms. Image quality assessment showed that ACT substantially improved image quality as compared to ECG (image quality score at end-diastole: ECG = 1.7 ± 0.5, ACT = 2.4 ± 0.5, p = 0.04) while the comparison between ECG vs. POX gated acquisitions showed no significant differences in image quality (image quality score: ECG = 1.7 ± 0.5, POX = 2.0 ± 0.5, p = 0.34). Conclusions The applicability of acoustic triggering for cardiac CINE imaging at 7.0 T was demonstrated. ACT's trigger reliability and fidelity are superior to that of ECG and POX. ACT promises to be beneficial for cardiovascular magnetic resonance at ultra-high field strengths including 7.0 T. KW - Interval Time Series KW - Image Quality Score KW - Image Quality Assessment KW - Sound Pressure Level KW - Cardiovascular Magnetic Resonance Y1 - 2010 U6 - https://doi.org/10.1186/1532-429X-12-67 SN - 1532-429X VL - 12 IS - 1 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Kob, Malte A1 - Frauenrath, Tobias T1 - A system for parallel measurement of glottis opening and larynx position JF - Biomedical Signal Processing and Control N2 - The simultaneous assessment of glottal dynamics and larynx position can be beneficial for the diagnosis of disordered voice or speech production and swallowing. Up to now, methods either concentrate on assessment of the glottis opening using optical, acoustical or electrical (electroglottography, EGG) methods, or on visualisation of the larynx position using ultrasound, computer tomography or magnetic resonance imaging techniques. The method presented here makes use of a time-multiplex measurement approach of space-resolved transfer impedances through the larynx. The fast sequence of measurements allows a quasi simultaneous assessment of both larynx position and EGG signal using up to 32 transmit–receive signal paths. The system assesses the dynamic opening status of the glottis as well as the vertical and back/forward motion of the larynx. Two electrode-arrays are used for the measurement of the electrical transfer impedance through the neck in different directions. From the acquired data the global and individual conductivity is calculated as well as a 2D point spatial representation of the minimum impedance. The position information is shown together with classical EGG signals allowing a synchronous visual assessment of glottal area and larynx position. A first application to singing voice analysis is presented that indicate a high potential of the method for use as a non-invasive tool in the diagnosis of voice, speech, and swallowing disorders. KW - EGG KW - Transfer impedance KW - Tomography KW - Larynx position KW - Voice assessment Y1 - 2009 U6 - https://doi.org/10.1016/j.bspc.2009.03.004 SN - 1746-8108 VL - 4 IS - 3 SP - 221 EP - 228 PB - Elsevier CY - Amsterdam ER - TY - JOUR A1 - Heinrichs, Uwe A1 - Utting, Jane F. A1 - Frauenrath, Tobias A1 - Hezel, Fabian A1 - Krombach, Gabriele A. A1 - Hodenius, Michael A. J. A1 - Kozerke, Sebastian A1 - Niendorf, Thoralf T1 - Myocardial T2 mapping free of distortion using susceptibility-weighted fast spin-echo imaging: A feasibility study at 1.5 T and 3.0 T JF - Magnetic Resonance in Medicine N2 - This study demonstrates the feasibility of applying free-breathing, cardiac-gated, susceptibility-weighted fast spin-echo imaging together with black blood preparation and navigator-gated respiratory motion compensation for anatomically accurate T₂ mapping of the heart. First, T₂ maps are presented for oil phantoms without and with respiratory motion emulation (T₂ = (22.1 ± 1.7) ms at 1.5 T and T₂ = (22.65 ± 0.89) ms at 3.0 T). T₂ relaxometry of a ferrofluid revealed relaxivities of R2 = (477.9 ± 17) mM⁻¹s⁻¹ and R2 = (449.6 ± 13) mM⁻¹s⁻¹ for UFLARE and multiecho gradient-echo imaging at 1.5 T. For inferoseptal myocardial regions mean T₂ values of 29.9 ± 6.6 ms (1.5 T) and 22.3 ± 4.8 ms (3.0 T) were estimated. For posterior myocardial areas close to the vena cava T₂-values of 24.0 ± 6.4 ms (1.5 T) and 15.4 ± 1.8 ms (3.0 T) were observed. The merits and limitations of the proposed approach are discussed and its implications for cardiac and vascular T₂-mapping are considered. Y1 - 2009 U6 - https://doi.org/10.1002/mrm.22054 SN - 1522-2594 VL - 62 IS - 3 SP - 822 EP - 828 PB - Wiley-Liss CY - New York ER - TY - JOUR A1 - Frauenrath, Tobias A1 - Niendorf, Thoralf A1 - Kob, Malte T1 - Acoustic method for synchronization of Magnetic Resonance Imaging (MRI) JF - Acta Acustica N2 - Magnetic Resonance Imaging (MRI) of moving organs requires synchronization with physiological motion or flow, which dictate the viable window for data acquisition. To meet this challenge, this study proposes an acoustic gating device (ACG) that employs acquisition and processing of acoustic signals for synchronization while providing MRI compatibility, immunity to interferences with electro-magnetic and acoustic fields and suitability for MRI at high magnetic field strengths. The applicability and robustness of the acoustic gating approach is examined in a pilot study, where it substitutes conventional ECG-gating for cardiovascular MR. The merits and limitations of the ACG approach are discussed. Implications for MR imaging in the presence of physiological motion are considered including synchronization with other structure- or motion borne sounds. Y1 - 2008 U6 - https://doi.org/10.3813/AAA.918017 SN - 1861-9959 VL - 94 IS - 1 SP - 148 EP - 155 PB - Hirzel CY - Stuttgart ER - TY - JOUR A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Frauenrath, Tobias A1 - Prothmann, Marcel A1 - Dieringer, Matthias A. A1 - Hezel, Fabian A1 - Renz, Wolfgang A1 - Kretschel, Kerstin A1 - Niendorf, Thoralf A1 - Schulz-Menger, Jeanette T1 - Cardiac chamber quantification using magnetic resonance imaging at 7 Tesla—a pilot study N2 - Objectives Interest in cardiovascular magnetic resonance (CMR) at 7 T is motivated by the expected increase in spatial and temporal resolution, but the method is technically challenging. We examined the feasibility of cardiac chamber quantification at 7 T. Methods A stack of short axes covering the left ventricle was obtained in nine healthy male volunteers. At 1.5 T, steady-state free precession (SSFP) and fast gradient echo (FGRE) cine imaging with 7 mm slice thickness (STH) were used. At 7 T, FGRE with 7 mm and 4 mm STH were applied. End-diastolic volume, end-systolic volume, ejection fraction and mass were calculated. Results All 7 T examinations provided excellent blood/myocardium contrast for all slice directions. No significant difference was found regarding ejection fraction and cardiac volumes between SSFP at 1.5 T and FGRE at 7 T, while volumes obtained from FGRE at 1.5 T were underestimated. Cardiac mass derived from FGRE at 1.5 and 7 T was larger than obtained from SSFP at 1.5 T. Agreement of volumes and mass between SSFP at 1.5 T and FGRE improved for FGRE at 7 T when combined with an STH reduction to 4 mm. Conclusions This pilot study demonstrates that cardiac chamber quantification at 7 T using FGRE is feasible and agrees closely with SSFP at 1.5 T. Y1 - 2010 U6 - https://doi.org/10.1007/s00330-010-1888-2 SN - 0938-7994 VL - 20 SP - 2844 EP - 2852 PB - Springer CY - Berlin, Heidelberg ER - TY - JOUR A1 - Droop, Philipp A1 - Chen, Shaohuang A1 - Radford, Melissa J. A1 - Paulßen, Elisabeth A1 - Gates, Byron D. A1 - Reilly, Raymond M. A1 - Radchenko, Valery A1 - Hoehr, Cornelia T1 - Synthesis of 197m/gHg labelled gold nanoparticles for targeted radionuclide therapy JF - Radiochimica Acta N2 - Meitner-Auger-electron emitters have a promising potential for targeted radionuclide therapy of cancer because of their short range and the high linear energy transfer of Meitner-Auger-electrons (MAE). One promising MAE candidate is 197m/gHg with its half-life of 23.8 h and 64.1 h, respectively, and high MAE yield. Gold nanoparticles (AuNPs) that are labelled with 197m/gHg could be a helpful tool for radiation treatment of glioblastoma multiforme when infused into the surgical cavity after resection to prevent recurrence. To produce such AuNPs, 197m/gHg was embedded into pristine AuNPs. Two different syntheses were tested starting from irradiated gold containing trace amounts of 197m/gHg. When sodium citrate was used as reducing agent, no 197m/gHg labelled AuNPs were formed, but with tannic acid, 197m/gHg labeled AuNPs were produced. The method was optimized by neutralizing the pH (pH = 7) of the Au/197m/gHg solution, which led to labelled AuNPs with a size of 12.3 ± 2.0 nm as measured by transmission electron microscopy. The labelled AuNPs had a concentration of 50 μg (gold)/mL with an activity of 151 ± 93 kBq/mL (197gHg, time corrected to the end of bombardment). KW - 197m/gHg KW - Gold nanoparticle (AuNP) KW - Meitner-Auger-electron (MAE) KW - Targeted radionuclide therapy (TRT) Y1 - 2023 U6 - https://doi.org/10.1515/ract-2023-0144 SN - 2193-3405 VL - 111 IS - 10 SP - 773 EP - 779 PB - De Gruyter CY - Berlin [u.a.] ER -