TY - JOUR A1 - Martin, Conrad A1 - Frauenrath, Tobias A1 - Özerdem, Celal A1 - Renz, Wolfgang A1 - Niendorf, Thoralf T1 - Development and evaluation of a small and mobile Magneto Alert Sensor (MALSE) to support safety requirements for magnetic resonance imaging JF - European Radiology N2 - Objective The purpose of this study is to (i) design a small and mobile Magnetic field ALert SEnsor (MALSE), (ii) to carefully evaluate its sensors to their consistency of activation/deactivation and sensitivity to magnetic fields, and (iii) to demonstrate the applicability of MALSE in 1.5 T, 3.0 T and 7.0 T MR fringe field environments. Methods MALSE comprises a set of reed sensors, which activate in response to their exposure to a magnetic field. The activation/deactivation of reed sensors was examined by moving them in/out of the fringe field generated by 7TMR. Results The consistency with which individual reed sensors would activate at the same field strength was found to be 100% for the setup used. All of the reed switches investigated required a substantial drop in ambient magnetic field strength before they deactivated. Conclusions MALSE is a simple concept for alerting MRI staff to a ferromagnetic object being brought into fringe magnetic fields which exceeds MALSEs activation magnetic field. MALSE can easily be attached to ferromagnetic objects within the vicinity of a scanner, thus creating a barrier for hazardous situations induced by ferromagnetic parts which should not enter the vicinity of an MR-system to occur. KW - MRI KW - MR safety KW - Magneto alert sensor KW - High field MRI KW - Uktrahigh field MRI Y1 - 2011 U6 - https://doi.org/10.1007/s00330-011-2153-z SN - 1432-1084 VL - 21 SP - 2187 EP - 2192 PB - Springer CY - Berlin, Heidelberg ER - TY - JOUR A1 - Dieringer, Matthias A. A1 - Renz, Wolfgang A1 - Lindel, Tomasz D. A1 - Seifert, Frank A1 - Frauenrath, Tobias A1 - von Knobelsdorf-Brenkenhoff, Florian A1 - Waiczies, Helmar A1 - Hoffmann, Werner A1 - Rieger, Jan A1 - Pfeiffer, Harald A1 - Ittermann, Bernd A1 - Schulz-Menger, Jeanette A1 - Niendorf, Thoralf T1 - Design and application of a four-channel transmit/receive surface coil for functional cardiac imaging at 7T JF - Journal of Magnetic Resonance Imaging N2 - Purpose To design and evaluate a four-channel cardiac transceiver coil array for functional cardiac imaging at 7T. Materials and Methods A four-element cardiac transceiver surface coil array was developed with two rectangular loops mounted on an anterior former and two rectangular loops on a posterior former. specific absorption rate (SAR) simulations were performed and a Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-1 calibration method was applied prior to obtain 2D FLASH CINE (mSENSE, R = 2) images from nine healthy volunteers with a spatial resolution of up to 1 × 1 × 2.5 mm3. Results Tuning and matching was found to be better than 10 dB for all subjects. The decoupling (S21) was measured to be >18 dB between neighboring loops, >20 dB for opposite loops, and >30 dB for other loop combinations. SAR values were well within the limits provided by the IEC. Imaging provided clinically acceptable signal homogeneity with an excellent blood-myocardium contrast applying the Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-2 calibration approach. Conclusion A four-channel cardiac transceiver coil array for 7T was built, allowing for cardiac imaging with clinically acceptable signal homogeneity and an excellent blood-myocardium contrast. Minor anatomic structures, such as pericardium, mitral, and tricuspid valves and their apparatus, as well as trabeculae, were accurately delineated. Y1 - 2011 U6 - https://doi.org/10.1002/jmri.22451 SN - 1522-2586 VL - 33 IS - 3 SP - 736 EP - 741 PB - Wiley-Liss CY - New York ER -