Design optimization of the PMT-ClearPET prototypes based on simulation studies with GEANT3

  • Within the Crystal Clear Collaboration four centres are developing 2nd generation high performance small animal PET scanners for different kinds of animals and medical applications. The first prototypes are PMT-based systems including depth of interaction (DOI) detection by using a phoswich layer of LSO and LuYAP. The aim of these simulation studies is to optimize sensitivity and spatial resolution of given designs, which vary in FOVs caused by different detector configurations (ring/octagon) and sizes. For this purpose the simulation tool GEANT3 (CERN) was used. The simulations have shown that all PMT designs with one-to-one coupling of crystals have a very nonlinear axial sensitivity profile. By shifting every other PMT 1/4 of a PMT length in axial direction the sampling of the FOVs became more homogeneous. At an energy threshold of 350keV the regression coefficient increases from 0.818 for the non-shifted to 0.993 for the shifted design. Simulations of a point source centred in the FOV (threshold: 350keV) resulted in sensitivities of 4.2% for a 4×20PMT (LSO/LuYAP a 10mm) and 3.8% for a 4×16PMT (LSO/LuYAP a 8mm) ring design. The 3D-MLEM reconstruction of a point source shows the enormous improvement of resolution using a crystal double layer with DOI (3.1mm at 40mm from CFOV) instead of a 20mm single layer (11.9mm).

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar
Metadaten
Verfasserangaben:U. Heinrichs, U. Pietrzyk, Karl Ziemons
ISSN:1082-3654
Titel des übergeordneten Werkes (Englisch):2002 IEEE Nuclear Science Symposium Conference Record, Vol. 3
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Erscheinungsjahr:2003
Datum der Publikation (Server):29.01.2013
Erste Seite:682
Letzte Seite:686
Link:http://dx.doi.org/10.1109/NSSMIC.2002.1239415
Zugriffsart:campus