• Treffer 2 von 85
Zurück zur Trefferliste

Robustness of neural networks algorithm for gamma detection in monolithic block detector, positron emission tomography

  • The monolithic scintillator block approach for gamma detection in the Positron Emission Tomography (PET) avoids estimating Depth of Interaction (DOI), reduces dead zones in detector and diminishes costs of detector production. Neural Networks (NN) are very efficient to determine the entrance point of a gamma incident on a scintillator block. This paper presents results on the robustness of the spatial resolution as a function of the random fraction in the data, temperature and HV fluctuations. This is important when implementing the method in a real scanner. Measurements were done with two Hamamatsu S8550 APD arrays, glued on a 20 × 20 × 10 mm3 monolithic LSO crystal block.

Metadaten exportieren

Weitere Dienste

Teilen auf X Suche bei Google Scholar
Metadaten
Verfasserangaben:M. Wedrowski, P. Bruyndonckx, S. Tavernier, L. Zhi, J. Dang, P. R. Mendes, J. M. Perez, Karl Ziemons
ISBN:1082-3654
Titel des übergeordneten Werkes (Englisch):2009 IEEE Nuclear Science Symposium Conference Record (NSS/MIC)
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Erscheinungsjahr:2009
Datum der Publikation (Server):04.02.2013
Erste Seite:2625
Letzte Seite:2628
Link:http://dx.doi.org/10.1109/NSSMIC.2009.5402007
Zugriffsart:campus
Fachbereiche und Einrichtungen:FH Aachen / Fachbereich Medizintechnik und Technomathematik