High speed and high resolution chemical imaging based on a new type of OLED-LAPS set-up

  • Light-addressable potentiometric sensors (LAPS) are field-effect-based sensors. A modulated light source is used to define the particular measurement spot to perform spatially resolved measurements of chemical species and to generate chemical images. In this work, an organic-LED (OLED) display has been chosen as a light source. This allows high measurement resolution and miniaturisation of the system. A new developed driving method for the OLED display optimised for LAPS-based measurements is demonstrated. The new method enables to define modulation frequencies between 1 kHz and 16 kHz and hence, reduces the measurement time of a chemical image by a factor of 40 compared to the traditional addressing of an OLED display.

Export metadata

Additional Services

Share in Twitter Search Google Scholar
Metadaten
Author:Frederik WernerORCiD, Torsten WagnerORCiD, Ko-ichiro Miyamoto, Tatsuo Yoshinobu, Michael Josef SchöningORCiD
DOI:https://doi.org/10.1016/j.snb.2011.12.102
ISSN:0925-4005
Parent Title (English):Sensors and Actuators B: Chemical
Publisher:Elsevier
Place of publication:Amsterdam
Document Type:Article
Language:English
Year of Completion:2012
Date of the Publication (Server):2012/12/18
Volume:175
First Page:118
Last Page:122
Note:
Part of special issue "Selected Papers presented at Eurosensors XXV"
Link:https://doi.org/10.1016/j.snb.2011.12.102
Zugriffsart:campus
Institutes:FH Aachen / Fachbereich Medizintechnik und Technomathematik
FH Aachen / INB - Institut für Nano- und Biotechnologien