• Treffer 33 von 63
Zurück zur Trefferliste

Pulse shape discrimination of LSO and LuYAP scintillators for depth of interaction detection in PET

  • A feasible way to gain the depth of interaction information in a positron emission tomography scanner is the use of phoswich detectors. In general, the layer of interaction is identified from the pulse shape of the corresponding scintillator material. In this work, pulses from LSO and LuYAP crystals were investigated in order to find a practical method of distinguishing. It turned out that such a pulse processing could be kept simple because of an additional slow component in the light decay of the LuYAP pulse. At the same time, the short decay time guarantees that the major amount of the light output is still collected within a short pulse recording time.

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar
Metadaten
Verfasserangaben:M. Streun, G. Brandenburg, H. Larue, H. Saleh, E. Zimmermann, Karl Ziemons, H. Halling
ISBN:0018-9499
Titel des übergeordneten Werkes (Englisch):IEEE Transactions on Nuclear Science
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Erscheinungsjahr:2003
Datum der Publikation (Server):29.01.2013
Jahrgang:50
Ausgabe / Heft:3
Erste Seite:344
Letzte Seite:347
Link:http://dx.doi.org/10.1109/TNS.2003.812449
Zugriffsart:campus
Fachbereiche und Einrichtungen:FH Aachen / Fachbereich Medizintechnik und Technomathematik