Refine
Year of publication
Institute
Has Fulltext
- no (205) (remove)
Document Type
- Article (171)
- Conference Proceeding (23)
- Part of a Book (10)
- Book (1)
Keywords
- Field-effect sensor (3)
- capacitive field-effect sensor (3)
- field-effect sensor (3)
- tobacco mosaic virus (TMV) (3)
- LAPS (2)
- gold nanoparticles (2)
- (Bio)degradation (1)
- Biomolecular logic gate (1)
- CNOT (1)
- Capacitive field-effect (1)
Electrolyte-insulator-semiconductor capacitors (EISCAP) belong to field-effect sensors having an attractive transducer architecture for constructing various biochemical sensors. In this study, a capacitive model of enzyme-modified EISCAPs has been developed and the impact of the surface coverage of immobilized enzymes on its capacitance-voltage and constant-capacitance characteristics was studied theoretically and experimentally. The used multicell arrangement enables a multiplexed electrochemical characterization of up to sixteen EISCAPs. Different enzyme coverages have been achieved by means of parallel electrical connection of bare and enzyme-covered single EISCAPs in diverse combinations. As predicted by the model, with increasing the enzyme coverage, both the shift of capacitance-voltage curves and the amplitude of the constant-capacitance signal increase, resulting in an enhancement of analyte sensitivity of the EISCAP biosensor. In addition, the capability of the multicell arrangement with multi-enzyme covered EISCAPs for sequentially detecting multianalytes (penicillin and urea) utilizing the enzymes penicillinase and urease has been experimentally demonstrated and discussed.