Refine
Year of publication
Document Type
- Article (20)
- Book (1)
- Conference Proceeding (1)
Keywords
- Raman spectroscopy (2)
- biosensors (2)
- capacitive field-effect sensors (2)
- Bacillus atrophaeus spores (1)
- DPA (dipicolinic acid) (1)
- acetoin (1)
- acetoin reductase (1)
- alcoholic beverages (1)
- atomic layer deposition (1)
- biocompatible materials (1)
- biodegradable electronic devices (1)
- capacitive electrolyte–insulator–semiconductor sensors (1)
- carbon electrodes (1)
- glucose (1)
- graphene oxide (1)
- hydrogen peroxide (1)
- layer-by-layer technique (1)
- nanomaterials (1)
- optical sensor setup (1)
- optical spore trapping (1)
An amperometric bi-enzyme biosensor based on substrate recycling principle for the amplification of the sensor signal has been developed for the detection of adrenaline in blood. Adrenaline can be used as biomarker verifying successful adrenal venous sampling procedure. The adrenaline biosensor has been realized via modification of a galvanic oxygen sensor with a bi-enzyme membrane combining a genetically modified laccase and a pyrroloquinoline quinone-dependent glucose dehydrogenase. The measurement conditions such as pH value and temperature were optimized to enhance the sensor performance. A high sensitivity and a low detection limit of about 0.5–1 nM adrenaline have been achieved in phosphate buffer at pH 7.4, relevant for measurements in blood samples. The sensitivity of the biosensor to other catecholamines such as noradrenaline, dopamine and dobutamine has been studied. Finally, the sensor has been successfully applied for the detection of adrenaline in human blood plasma.