To design and evaluate a four-channel cardiac transceiver coil array for functional cardiac imaging at 7T.
Materials and Methods
A four-element cardiac transceiver surface coil array was developed with two rectangular loops mounted on an anterior former and two rectangular loops on a posterior former. specific absorption rate (SAR) simulations were performed and a Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-1 calibration method was applied prior to obtain 2D FLASH CINE (mSENSE, R = 2) images from nine healthy volunteers with a spatial resolution of up to 1 × 1 × 2.5 mm3.
Results
Tuning and matching was found to be better than 10 dB for all subjects. The decoupling (S21) was measured to be >18 dB between neighboring loops, >20 dB for opposite loops, and >30 dB for other loop combinations. SAR values were well within the limits provided by the IEC. Imaging provided clinically acceptable signal homogeneity with an excellent blood-myocardium contrast applying the Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-2 calibration approach.
Conclusion
A four-channel cardiac transceiver coil array for 7T was built, allowing for cardiac imaging with clinically acceptable signal homogeneity and an excellent blood-myocardium contrast. Minor anatomic structures, such as pericardium, mitral, and tricuspid valves and their apparatus, as well as trabeculae, were accurately delineated.
4CH TX/RX Surface Coil for 7T: Design, Optimization and Application for Cardiac Function Imaging
(2010)
Practical impediments of ultra high field cardiovascular MR (CVMR) can be catalogued in exacerbated magnetic field and radio frequency (RF) inhomogeneities, susceptibility and off-resonance effects, conductive and dielectric effects in tissue, and RF power deposition constraints, which all bear the potential to spoil the benefit of CVMR at 7T. Therefore, a four element cardiac transceive surface coil array was developed. Cardiac imaging provided clinically acceptable signal homogeneity with an excellent blood myocardium contrast. Subtle anatomic structures, such as pericardium, mitral and tricuspid valves and their apparatus, papillary muscles, and trabecles were accurately delineated.