As ultrahigh field cardiac MRI becomes more widespread in the (pre)clinical research arena the propensity of ECG recordings to interference from electromagnetic fields and magneto-hydrodynamic effects increases and with it the motivation for a practical gating/triggering alternative. This study compares the feasibility, efficacy and reliability phonocardiogram (ACT), vector electrocardiogram (VCG) and traditional pulse oximetry (pO2) triggered MRI for left ventricular function assessment at 7.0T. ACTs intrinsic insensitivity to interference from electro-magnetic fields and magneto-hydrodynamic effects results in an excellent trigger reliability and renders it suitable for global cardiac function assessment at ultrahigh magnetic field strengths.
ECG is corrupted by magneto-hydrodynamic effects at higher magnetic field strength. Artifacts in the ECG trace and severe T-wave elevation might be mis-interpreted as R-waves. MHD being inherently sensitive to blood flow and blood velocity provides an alternative approach for cardiac gating, even in peripheral target areas far away from the commonly used upper torso positions of ECG electrodes. This feature would be very beneficial to address traveling time induced motion artifacts and trigger latency related issues raised by ECG-gated peripheral MR angiography. For all those reasons, this work proposes the use of MHD-trigger for cardiac gated MR.
4CH TX/RX Surface Coil for 7T: Design, Optimization and Application for Cardiac Function Imaging
(2010)
Practical impediments of ultra high field cardiovascular MR (CVMR) can be catalogued in exacerbated magnetic field and radio frequency (RF) inhomogeneities, susceptibility and off-resonance effects, conductive and dielectric effects in tissue, and RF power deposition constraints, which all bear the potential to spoil the benefit of CVMR at 7T. Therefore, a four element cardiac transceive surface coil array was developed. Cardiac imaging provided clinically acceptable signal homogeneity with an excellent blood myocardium contrast. Subtle anatomic structures, such as pericardium, mitral and tricuspid valves and their apparatus, papillary muscles, and trabecles were accurately delineated.
With its need for high SNR and short acquisition times, Cardiac MRI (CMR) is an intriguing target application for ultrahigh field MRI. Due to the sheer size of the upper torso, however, the known RF issues of 7T MRI are also most prominent in CMR. Recent years brought substantial progress but the full potential of the ultrahigh field for CMR is yet to be exploited. Parallel transmission (pTx) is a promising approach in this context and several groups have already reported B1 shimming for 7T CMR. In such a static pTx application amplitudes and phases of all Tx channels are adjusted individually but otherwise imaging techniques established in current clinical practice 1.5 T and 3 T are applied. More advanced forms of pTx as spatially selective excitation (SSE) using Transmit SENSE promise additional benefits like faster imaging with reduced fields of view or improved SAR control. SSE requires the full dynamic capabilities of pTx, however, and for the majority of today's implemented pTx hardware the internal synchronization of the Tx array does not easily permit external triggering as needed for CMR. Here we report a software solution to this problem and demonstrate the feasibility of CINE CMR at 7 T using a Tx array.
Cardiac MR (CMR) is of proven clinical value but also an area of vigorous ongoing research since image quality is not always exclusively defined by signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Recent developments of CMR at 7.0 T have been driven by pioneering explorations into novel multichannel transmit and receive coil array technology to tackle the challenges B1+-field inhomogeneities, to offset specific-absorption rate (SAR) constraints and to reduce banding artifacts in SSFP imaging. For this study, recognition of the benefits and performance of local surface Tx/Rx-array structures recently established at 7.0 T inspired migration to 3.0 T, where RF inhomogeneities and SAR limitations encountered in routine clinical CMR, though somewhat reduced versus the 7.0 T situation, remain significant. For all these reasons, this study was designed to build and examine the feasibility of a local four channel Tx/Rx cardiac coil array for anatomical and functional cardiac imaging at 3.0 T. For comparison, a homebuilt 4 channel Rx cardiac coil array exhibiting the same geometry as the Tx/Rx coil and a Rx surface coil array were used.
In current clinical cardiovascular MR (CMR) practice cardiac motion is commonly dealt with using ECG based synchronization. However, ECG is corrupted by magneto-hydrodynamic (MHD) effects in magnetic fields. This leads to artifacts in the ECG trace and evokes severe T-wave elevations, which might be misinterpreted as R-waves resulting in erroneous triggering. At (ultra)high field strengths, the propensity of ECG recordings to MHD effects is further pronounced. Pulse oximetry (POX) being inherently sensitive to blood oxygenation provides an alternative approach for cardiac gating. However, due to the travel time of the blood the peak of maximum oxygenation and hence the trigger is delayed by approx. 300 ms with respect to the ECG's R-wave. Also the peak of maximum oxygenation shows a jitter of up to 65 ms. Alternative triggering approaches include acoustic cardiac triggering (ACT). In current clinical practice cardiac gating / triggering commonly relies on using single physiological signals only. Realizing this limitation this study proposes a combined triggering approach which exploits multiple physiological signals including ECG, POX or ACT to track cardiac activity. The feasibility of the coupled approach is examined for LV function assessment at 7.0 T. For this purpose, breath-held 2D-CINE imaging in conjunction with cardiac synchronization was performed paralleled by real time logging of physiological waveforms to track (mis)synchronization between the cardiac cycle and data acquisition. Combinations of the ECG, POX and ACT signals were evaluated and processed in real time to facilitate reliable trigger information.
The assessment of the right ventricle (RV) is a challenge in today's cardiology, but of growing clinical impact regarding patient prognosis in different cardiac diseases. The detection and differentiation of small wall motion abnormalities may help to enhance the differentiation of cardiomyopathies including Arrhythmogenic Rightventricular Cardiomyopathy. Cardiovascular magnetic resonance (CMR) at 1.5T is the accepted gold standard for RV quantification. The higher spatial resolution achievable at ultrahigh field strength (UHF) offers the potential to gain new insights into the structure and function of the RV. To approach this goal accurate RV chamber quantification at 7T has to be proven. Consequently this study examines the feasibility of assessment of RV dimensions and function at 7T using improved spatial resolution enabled by the intrinsic sensitivity gain of UHF CMR. For this purpose, a dedicated 16 channel TX/RX RF coil array is used together with 2D CINE fast gradient echo (FGRE) imaging. For comparison RV chamber quantification is conducted at 1.5T using a SSFP based state of the art clinical protocol.
Objectives
Interest in cardiovascular magnetic resonance (CMR) at 7 T is motivated by the expected increase in spatial and temporal resolution, but the method is technically challenging. We examined the feasibility of cardiac chamber quantification at 7 T.
Methods
A stack of short axes covering the left ventricle was obtained in nine healthy male volunteers. At 1.5 T, steady-state free precession (SSFP) and fast gradient echo (FGRE) cine imaging with 7 mm slice thickness (STH) were used. At 7 T, FGRE with 7 mm and 4 mm STH were applied. End-diastolic volume, end-systolic volume, ejection fraction and mass were calculated.
Results
All 7 T examinations provided excellent blood/myocardium contrast for all slice directions. No significant difference was found regarding ejection fraction and cardiac volumes between SSFP at 1.5 T and FGRE at 7 T, while volumes obtained from FGRE at 1.5 T were underestimated. Cardiac mass derived from FGRE at 1.5 and 7 T was larger than obtained from SSFP at 1.5 T. Agreement of volumes and mass between SSFP at 1.5 T and FGRE improved for FGRE at 7 T when combined with an STH reduction to 4 mm.
Conclusions
This pilot study demonstrates that cardiac chamber quantification at 7 T using FGRE is feasible and agrees closely with SSFP at 1.5 T.
Purpose
To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T.
Methods
The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study.
Results
Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly.
Conclusion
The modular 32-channel transceiver cardiac array supports accelerated and high spatial resolution cardiac MRI. The array is compatible with multichannel transmission and provides a technological basis for future clinical assessment of parallel transmission techniques at 7.0T.