Detection of hydrogen peroxide vapor by use of manganese(IV) oxide as catalyst for calorimetric gas sensors
- In this work, the catalyst manganese(IV) oxide (MnO2), of calorimetric gas sensors (to monitor the sterilization agent vaporized hydrogen peroxide) has been investigated in more detail. Chemical analyses by means of X-ray-induced photoelectron spectroscopy have been performed to unravel the surface chemistry prior and after exposure to hydrogen peroxide vapor at elevated temperature, as applied in the sterilization processes of beverage cartons. The surface characterization reveals a change in oxidation states of the metal oxide catalyst after exposure to hydrogen peroxide. Additionally, a cleaning effect of the catalyst, which itself is attached to the sensor surface by means of a polymer interlayer, could be observed.
Author: | Jan OberländerORCiD, Patrick Kirchner, Hans-Gerd Boyen, Michael Josef SchöningORCiD |
---|---|
DOI: | https://doi.org/10.1002/pssa.201330359 |
ISSN: | 1521-396X (E-Journal); 1862-6319 (E-Journal); 0031-8965 (Print); 1862-6300 (Print) |
Parent Title (English): | Physica status solidi A: Applications and materials science |
Publisher: | Wiley-VCH |
Place of publication: | Weinheim |
Document Type: | Article |
Language: | English |
Year of Completion: | 2014 |
Volume: | 211 |
Issue: | 6 |
First Page: | 1372 |
Last Page: | 1376 |
Link: | https://doi.org/10.1002/pssa.201330359 |
Zugriffsart: | campus |
Institutes: | FH Aachen / Fachbereich Medizintechnik und Technomathematik |
FH Aachen / INB - Institut für Nano- und Biotechnologien | |
collections: | Verlag / Wiley-VCH |