Refine
Year of publication
Document Type
- Article (47)
- Conference Proceeding (5)
- Book (1)
- Part of a Book (1)
Keywords
- Light-addressable potentiometric sensor (3)
- LAPS (2)
- Chemical images (1)
- Chemical imaging (1)
- Chemical imaging sensor (1)
- Chemical sensor (1)
- EIS capacitive sensor (1)
- Field-effect device (1)
- Light-addressable Potentiometric Sensor (1)
- MOS (1)
- Negative impedance convertor (1)
- Organic light-emitting diode display (1)
- Potentiometry (1)
- Resonance-mode measurement (1)
- artificial olfactory image (1)
- catalytic metal (1)
- chemical sensor (1)
- electronic nose (1)
- field-effect structure (1)
- gas sensor (1)
The light-addressable potentiometric sensor (LAPS) has the unique feature to address different regions of a sensor surface without the need of complex structures. Measurements at different locations on the sensor surface can be performed in a common analyte solution, which distinctly simplifies the fluidic set-up. However, the measurement in a single analyte chamber prevents the application of different drugs or different concentrations of a drug to each measurement spot at the same time as in the case of multi-reservoir-based set-ups. In this work, the authors designed a LAPS-based set-up for cell culture screening that utilises magnetic beads loaded with the endotoxin (lipopolysaccharides, LPS), to generate a spatially distributed gradient of analyte concentration. Different external magnetic fields can be adjusted to move the magnetic beads loaded with a specific drug within the measurement cell. By recording the metabolic activities of a cell layer cultured on top of the LAPS surface, this work shows the possibility to apply different concentrations of a sample along the LAPS measurement spots within a common analyte solution.