In this paper we present an extension of the action language Golog that allows for using fuzzy notions in non-deterministic argument choices and the reward function in decision-theoretic planning. Often, in decision-theoretic planning, it is cumbersome to specify the set of values to pick from in the non-deterministic-choice-of-argument statement. Also, even for domain experts, it is not always easy to specify a reward function. Instead of providing a finite domain for values in the non-deterministic-choice-of-argument statement in Golog, we now allow for stating the argument domain by simply providing a formula over linguistic terms and fuzzy uents. In Golog’s forward-search DT planning algorithm, these formulas are evaluated in order to find the agent’s optimal policy. We illustrate this in the Diner Domain where the agent needs to calculate the optimal serving order.
Purpose
To design and evaluate a four-channel cardiac transceiver coil array for functional cardiac imaging at 7T.
Materials and Methods
A four-element cardiac transceiver surface coil array was developed with two rectangular loops mounted on an anterior former and two rectangular loops on a posterior former. specific absorption rate (SAR) simulations were performed and a Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-1 calibration method was applied prior to obtain 2D FLASH CINE (mSENSE, R = 2) images from nine healthy volunteers with a spatial resolution of up to 1 × 1 × 2.5 mm3.
Results
Tuning and matching was found to be better than 10 dB for all subjects. The decoupling (S21) was measured to be >18 dB between neighboring loops, >20 dB for opposite loops, and >30 dB for other loop combinations. SAR values were well within the limits provided by the IEC. Imaging provided clinically acceptable signal homogeneity with an excellent blood-myocardium contrast applying the Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-2 calibration approach.
Conclusion
A four-channel cardiac transceiver coil array for 7T was built, allowing for cardiac imaging with clinically acceptable signal homogeneity and an excellent blood-myocardium contrast. Minor anatomic structures, such as pericardium, mitral, and tricuspid valves and their apparatus, as well as trabeculae, were accurately delineated.
Design of a Measurement Setup for the Impedance of Low-Resistance Plate-Conductors in the RF-Range
(2006)
Purpose:
To investigate the feasibility of using magnetohydrodynamic (MHD) effects for synchronization of magnetic resonance imaging (MRI) with the cardiac cycle.
Materials and Methods:
The MHD effect was scrutinized using a pulsatile flow phantom at B0 = 7.0 T. MHD effects were examined in vivo in healthy volunteers (n = 10) for B0 ranging from 0.05–7.0 T. Noncontrast-enhanced MR angiography (MRA) of the carotids was performed using a gated steady-state free-precession (SSFP) imaging technique in conjunction with electrocardiogram (ECG) and MHD synchronization.
Results:
The MHD potential correlates with flow velocities derived from phase contrast MRI. MHD voltages depend on the orientation between B0 and the flow of a conductive fluid. An increase in the interelectrode spacing along the flow increases the MHD potential. In vivo measurement of the MHD effect provides peak voltages of 1.5 mV for surface areas close to the common carotid artery at B0 = 7.0 T. Synchronization of MRI with the cardiac cycle using MHD triggering is feasible. MHD triggered MRA of the carotids at 3.0 T showed an overall image quality and richness of anatomic detail, which is comparable to ECG-triggered MRAs.
Conclusion:
This feasibility study demonstrates the use of MHD effects for synchronization of MR acquisitions with the cardiac cycle. J. Magn. Reson. Imaging 2012;36:364–372. © 2012 Wiley Periodicals, Inc.
A novel method to determine the extruded length of a metallic wire for a directed energy deposition (DED) process using a microwave (MW) plasma jet with a straight-through wire feed is presented. The method is based on the relative comparison of the measured frequency response obtained by the large-signal scattering parameter (Hot-S) technique. In the practical working range, repeatability of less than 6% for a nonactive plasma and 9% for the active plasma state is found. Measurements are conducted with a focus on a simple solution to decrease the processing time and reduce the integration time of the process into the existing hardware. It is shown that monitoring a single frequency for magnitude and phase changes is sufficient to achieve good accuracy. A combination of different measurement values to determine the length is possible. The applicability to different diameter of the same material is shown as well as a contact detection of the wire and metallic substrate.
Objective
The purpose of this study is to (i) design a small and mobile Magnetic field ALert SEnsor (MALSE), (ii) to carefully evaluate its sensors to their consistency of activation/deactivation and sensitivity to magnetic fields, and (iii) to demonstrate the applicability of MALSE in 1.5 T, 3.0 T and 7.0 T MR fringe field environments.
Methods
MALSE comprises a set of reed sensors, which activate in response to their exposure to a magnetic field. The activation/deactivation of reed sensors was examined by moving them in/out of the fringe field generated by 7TMR.
Results
The consistency with which individual reed sensors would activate at the same field strength was found to be 100% for the setup used. All of the reed switches investigated required a substantial drop in ambient magnetic field strength before they deactivated.
Conclusions
MALSE is a simple concept for alerting MRI staff to a ferromagnetic object being brought into fringe magnetic fields which exceeds MALSEs activation magnetic field. MALSE can easily be attached to ferromagnetic objects within the vicinity of a scanner, thus creating a barrier for hazardous situations induced by ferromagnetic parts which should not enter the vicinity of an MR-system to occur.