Analysis of the long-term effect of the MBST® nuclear magnetic resonance therapy on gonarthrosis
(2016)
Introduction
In regard of surgical training, the reproducible simulation of life-like proximal humerus fractures in human cadaveric specimens is desirable. The aim of the present study was to develop a technique that allows simulation of realistic proximal humerus fractures and to analyse the influence of rotator cuff preload on the generated lesions in regards of fracture configuration.
Materials and methods
Ten cadaveric specimens (6 left, 4 right) were fractured using a custom-made drop-test bench, in two groups. Five specimens were fractured without rotator cuff preload, while the other five were fractured with the tendons of the rotator cuff preloaded with 2 kg each. The humeral shaft and the shortened scapula were potted. The humerus was positioned at 90° of abduction and 10° of internal rotation to simulate a fall on the elevated arm. In two specimens of each group, the emergence of the fractures was documented with high-speed video imaging. Pre-fracture radiographs were taken to evaluate the deltoid-tuberosity index as a measure of bone density. Post-fracture X-rays and CT scans were performed to define the exact fracture configurations. Neer’s classification was used to analyse the fractures.
Results
In all ten cadaveric specimens life-like proximal humerus fractures were achieved. Two III-part and three IV-part fractures resulted in each group. The preloading of the rotator cuff muscles had no further influence on the fracture configuration. High-speed videos of the fracture simulation revealed identical fracture mechanisms for both groups. We observed a two-step fracture mechanism, with initial impaction of the head segment against the glenoid followed by fracturing of the head and the tuberosities and then with further impaction of the shaft against the acromion, which lead to separation of the tuberosities.
Conclusion
A high energetic axial impulse can reliably induce realistic proximal humerus fractures in cadaveric specimens. The preload of the rotator cuff muscles had no influence on initial fracture configuration. Therefore, fracture simulation in the proximal humerus is less elaborate. Using the presented technique, pre-fractured specimens are available for real-life surgical education.
Autonomous robotic systems for penetrating thick ice shells with simultaneous collecting of scientific data are very promising devices in both terrestrial (glacier, climate research) and extra-terrestrial applications. Technical challenges in development of such systems are numerous and include 3D-navigation, an appropriate energy source, motion control, etc. Not less important is the problem of forward contamination of the pristine glacial environments with microorganisms and biomolecules from the surface of the probe. This study was devoted to establishing a laboratory model for microbial contamination of a newly constructed ice-melting probe called IceMole and to analyse the viability and amount of the contaminating microorganisms as a function of distance. The used bacterial strains were Bacillus subtilis (ATCC 6051) and Escherichia coli (ATCC 11775). The main objective was development of an efficient and reliable in-situ decontamination method of the melting probe. Therefore, several chemical substances were tested in respect of their efficacy to eliminate bacteria on the surface of the melting probe at low temperature (0 - 5 °C) and at continuous dilution by melted water. Our study has shown that at least 99.9% decontamination of the IceMole can be successfully achieved by the injection of 30% (v/v) hydrogen peroxide and 3% (v/v) sodium hypochlorite into the drilling site. We were able to reproduce this result in both time-dependent and depth-dependent experiments. The sufficient amount of 30% (v/v) H₂O₂ or 3% (v/v) NaClO has been found to be approximately 18 L per cm² of the probe’s surface.
False spectra formation in the differential two-channel scheme of the laser Doppler flowmeter
(2018)
Noise in the differential two-channel scheme of a classic laser Doppler flowmetry (LDF) instrument was studied. Formation of false spectral components in the output signal due to beating of electrical signals in the differential amplifier was found out. The improved block-diagram of the flowmeter was developed allowing to reduce the noise.
Bilaterale stromale Hornhauttrübungen sind für den Augenarzt eine differenzialdiagnostische Herausforderung. Im folgenden Beitrag werden 2 Patieninnen (30 und 36 Jahre) mit unterschiedlich stark ausgeprägter stromaler diffuser Hornhauttrübung vorgestellt. Patientin 1 war kleinwüchsig (114 cm) und Patientin 2 normal groß (172 cm). Beide Patientinnen wiesen veränderte Gelenkstrukturen an Hand und Fußgelenken sowie diffuse stromale Hornhauttrübungen auf. Des Weiteren lagen eine Mitral- und Aorteninsuffizienz (Patientin 1) bzw. eine Aorteninsuffizienz (Patientin 2) vor. Die stromalen diffusen Hornhauttrübungen ließen im Zusammenhang mit den Gelenkveränderungen ein Scheie-Syndrom vermuten. Therapeutisch ist bei Patienten mit Visusminderung eine (lamelläre) Keratoplastik sinnvoll.