We have developed a double-tuned ¹H/¹⁹F birdcage resonator dedicated for hand and wrist imaging at 7 T to locally image non-steroidal anti-inflammatory drugs (NSAID) such as 2-{[3-(Trifluoromethyl) phenyl]amino}benzoic acid. The preliminary in vivo images acquired by the double-tuned ¹H/¹⁹F birdcage resonator demonstrate the feasibility for ¹H/¹⁹F hand- and wrist-imaging at 7 T. While the diagnostic quality of the coil needs to be assessed in patients with inflammatory rheumatoid disease, first ¹⁹F images of the NSAID are encouraging, and point towards the prospect of applying ¹⁹F-MRI to visualize and quantify the concentration of therapeutically-active compound at the sites of inflammation.
Purpose
To design and evaluate a modular transceiver coil array with 32 independent channels for cardiac MRI at 7.0T.
Methods
The modular coil array comprises eight independent building blocks, each containing four transceiver loop elements. Numerical simulations were used for B1+ field homogenization and radiofrequency (RF) safety validation. RF characteristics were examined in a phantom study. The array's suitability for accelerated high spatial resolution two-dimensional (2D) FLASH CINE imaging of the heart was examined in a volunteer study.
Results
Transmission field adjustments and RF characteristics were found to be suitable for the volunteer study. The signal-to-noise intrinsic to 7.0T together with the coil performance afforded a spatial resolution of 1.1 × 1.1 × 2.5 mm3 for 2D CINE FLASH MRI, which is by a factor of 6 superior to standardized CINE protocols used in clinical practice at 1.5T. The 32-channel transceiver array supports one-dimensional acceleration factors of up to R = 4 without impairing image quality significantly.