Article
Refine
Year of publication
- 2012 (40) (remove)
Institute
- Fachbereich Medizintechnik und Technomathematik (40) (remove)
Has Fulltext
- no (40) (remove)
Document Type
- Article (40) (remove)
Keywords
- (Bio)degradation (1)
- Acceleration (1)
- Afterload (1)
- Anastomotic leakage (1)
- Autolysis (1)
- Calorimetric gas sensor (1)
- Cell permeability (1)
- CellDrum (1)
- Cellular force (1)
- Chemical imaging sensor (1)
Light-addressable potentiometric sensors (LAPS) are field-effect-based sensors. A modulated light source is used to define the particular measurement spot to perform spatially resolved measurements of chemical species and to generate chemical images. In this work, an organic-LED (OLED) display has been chosen as a light source. This allows high measurement resolution and miniaturisation of the system. A new developed driving method for the OLED display optimised for LAPS-based measurements is demonstrated. The new method enables to define modulation frequencies between 1 kHz and 16 kHz and hence, reduces the measurement time of a chemical image by a factor of 40 compared to the traditional addressing of an OLED display.