Refine
Year of publication
- 2024 (21)
- 2023 (22)
- 2022 (42)
- 2021 (43)
- 2020 (57)
- 2019 (64)
- 2018 (59)
- 2017 (60)
- 2016 (41)
- 2015 (59)
- 2014 (52)
- 2013 (53)
- 2012 (58)
- 2011 (65)
- 2010 (58)
- 2009 (66)
- 2008 (50)
- 2007 (40)
- 2006 (37)
- 2005 (36)
- 2004 (68)
- 2003 (38)
- 2002 (44)
- 2001 (46)
- 2000 (47)
- 1999 (29)
- 1998 (24)
- 1997 (22)
- 1996 (21)
- 1995 (16)
- 1994 (11)
- 1993 (16)
- 1992 (7)
- 1991 (5)
- 1990 (11)
- 1989 (10)
- 1988 (16)
- 1987 (6)
- 1986 (2)
- 1985 (2)
- 1984 (1)
- 1983 (2)
- 1982 (20)
- 1981 (13)
- 1980 (27)
- 1979 (18)
- 1978 (26)
- 1977 (13)
- 1976 (12)
- 1975 (9)
- 1974 (2)
- 1973 (1)
- 1972 (2)
- 1968 (1)
Institute
- Fachbereich Medizintechnik und Technomathematik (1571) (remove)
Has Fulltext
- no (1571) (remove)
Language
- English (1571) (remove)
Document Type
- Article (1316)
- Conference Proceeding (135)
- Book (43)
- Part of a Book (43)
- Doctoral Thesis (18)
- Other (6)
- Patent (4)
- Preprint (3)
- Conference: Meeting Abstract (1)
- Habilitation (1)
Keywords
- LAPS (4)
- Natural language processing (4)
- CellDrum (3)
- Field-effect sensor (3)
- Light-addressable potentiometric sensor (3)
- Paired sample (3)
- hydrogen peroxide (3)
- impedance spectroscopy (3)
- Bacillus atrophaeus (2)
- Biocomposites (2)
Impedance spectroscopy: A tool for real-time in situ monitoring of the degradation of biopolymers
(2013)
Investigation of the degradation kinetics of biodegradable polymers is essential for the development of implantable biomedical devices with predicted biodegradability. In this work, an impedimetric sensor has been applied for real-time and in situ monitoring of degradation processes of biopolymers. The sensor consists of two platinum thin-film electrodes covered by a polymer film to be studied. The benchmark biomedical polymer poly(D,L-lactic acid) (PDLLA) was used as a model system. PDLLA films were deposited on the sensor structure from a polymer solution by using the spin-coating method. The degradation kinetics of PDLLA films have been studied in alkaline solutions of pH 9 and 12 by means of an impedance spectroscopy (IS) method. Any changes in a polymer capacitance/resistance induced by water uptake and/or polymer degradation will modulate the global impedance of the polymer-covered sensor that can be used as an indicator of the polymer degradation. The degradation rate can be evaluated from the time-dependent impedance spectra. As expected, a faster degradation has been observed for PDLLA films exposed to pH 12 solution.
Designing novel or optimizing existing biodegradable polymers for biomedical applications requires numerous tests on the effect of substances on the degradation process. In the present work, polymer-modified electrolyte–insulator–semiconductor (PMEIS) sensors have been applied for monitoring an enzymatically catalyzed degradation of polymers for the first time. The thin films of biodegradable polymer poly(d,l-lactic acid) and enzyme lipase were used as a model system. During degradation, the sensors were read-out by means of impedance spectroscopy. In order to interpret the data obtained from impedance measurements, an electrical equivalent circuit model was developed. In addition, morphological investigations of the polymer surface have been performed by means of in situ atomic force microscopy. The sensor signal change, which reflects the progress of degradation, indicates an accelerated degradation in the presence of the enzyme compared to hydrolysis in neutral pH buffer media. The degradation rate increases with increasing enzyme concentration. The obtained results demonstrate the potential of PMEIS sensors as a very promising tool for in situ and real-time monitoring of degradation of polymers.
A sensor system for investigating (bio)degradationprocesses of polymers is presented. The system utilizes semiconductor field-effect sensors and is capable of monitoring the degradation process in-situ and in real-time. The degradation of the polymer poly(d,l-lactic acid) is exemplarily monitored in solutions with different pH value, pH-buffer solution containing the model enzyme lipase from Rhizomucormiehei and cell-culture medium containing supernatants from stimulated and non-stimulated THP-1-derived macrophages mimicking activation of the immune system.
A semiconductor field-effect device has been used for an enzymatically catalyzed degradation of biopolymers for the first time. This novel technique is capable to monitor the degradation process of multiple samples in situ and in real-time. As model system, the degradation of the biopolymer poly(D, L-lactic acid) has been monitored in the degradation medium containing the enzyme lipase from Rhizomucor miehei. The obtained results demonstrate the potential of capacitive field-effect sensors for degradation studies of biodegradable polymers.
Characterising an insect antenna as a receptor for a biosensor by means of impedance spectroscopy
(2001)
This paper presents NLP Lean Programming
framework (NLPf), a new framework
for creating custom natural language processing
(NLP) models and pipelines by utilizing
common software development build systems.
This approach allows developers to train and
integrate domain-specific NLP pipelines into
their applications seamlessly. Additionally,
NLPf provides an annotation tool which improves
the annotation process significantly by
providing a well-designed GUI and sophisticated
way of using input devices. Due to
NLPf’s properties developers and domain experts
are able to build domain-specific NLP
applications more efficiently. NLPf is Opensource
software and available at https://
gitlab.com/schrieveslaach/NLPf.
In this paper, carbon nanotubes (CNTs) were incorporated in penicillinase-phospholipid Langmuir and Langmuir–Blodgett (LB) films to enhance the enzyme catalytic properties. Adsorption of the penicillinase and CNTs at dimyristoylphosphatidic acid (DMPA) monolayers at the air–water interface was investigated by surface pressure–area isotherms, vibrational spectroscopy, and Brewster angle microscopy. The floating monolayers were transferred to solid supports through the LB technique, forming mixed DMPA-CNTs-PEN films, which were investigated by quartz crystal microbalance, vibrational spectroscopy, and atomic force microscopy. Enzyme activity was studied with UV–vis spectroscopy and the feasibility of the supramolecular device nanostructured as ultrathin films were essayed in a capacitive electrolyte–insulator–semiconductor (EIS) sensor device. The presence of CNTs in the enzyme–lipid LB film not only tuned the catalytic activity of penicillinase but also helped conserve its enzyme activity after weeks, showing increased values of activity. Viability as penicillin sensor was demonstrated with capacitance/voltage and constant capacitance measurements, exhibiting regular and distinctive output signals over all concentrations used in this work. These results may be related not only to the nanostructured system provided by the film, but also to the synergism between the compounds on the active layer, leading to a surface morphology that allowed a fast analyte diffusion because of an adequate molecular accommodation, which also preserved the penicillinase activity. This work therefore demonstrates the feasibility of employing LB films composed of lipids, CNTs, and enzymes as EIS devices for biosensing applications.