Refine
Year of publication
- 2011 (328) (remove)
Institute
- Fachbereich Medizintechnik und Technomathematik (82)
- Fachbereich Elektrotechnik und Informationstechnik (40)
- INB - Institut für Nano- und Biotechnologien (37)
- Fachbereich Wirtschaftswissenschaften (34)
- IfB - Institut für Bioengineering (34)
- Fachbereich Energietechnik (29)
- Fachbereich Chemie und Biotechnologie (28)
- Fachbereich Bauingenieurwesen (24)
- Fachbereich Maschinenbau und Mechatronik (24)
- Fachbereich Luft- und Raumfahrttechnik (23)
Document Type
- Article (182)
- Conference Proceeding (76)
- Book (24)
- Part of a Book (21)
- Report (7)
- Patent (5)
- Conference: Meeting Abstract (3)
- Other (3)
- Doctoral Thesis (2)
- Part of a Periodical (2)
Keywords
- Aachen / Fachhochschule Aachen (3)
- Aachen University of Applied Sciences (3)
- Aktionskunst (3)
- FH Aachen (2)
- Illustration (2)
- Interaktive Kunst (2)
- Kommunikationsguerrilla (2)
- Pflanzenphysiologie (2)
- Pflanzenscanner (2)
- Stadtgestaltung (2)
We present the novel concept of a combined drilling and melting probe for subsurface ice research. This probe, named “IceMole”, is currently developed, built, and tested at the FH Aachen University of Applied Sciences’ Astronautical Laboratory. Here, we describe its first prototype design and report the results of its field tests on the Swiss Morteratsch glacier. Although the IceMole design is currently adapted to terrestrial glaciers and ice shields, it may later be modified for the subsurface in-situ investigation of extraterrestrial ice, e.g., on Mars, Europa, and Enceladus. If life exists on those bodies, it may be present in the ice (as life can also be found in the deep ice of Earth).
Möglichkeiten und Grenzen der Anwendbarkeit statisch nichtlinearer Verfahren nach DIN EN 1998-1
(2011)
We present the novel concept of a combined drilling and melting probe for subsurface ice research. This probe, named “IceMole”, is currently developed, built, and tested at the FH Aachen University of Applied Sciences’ Astronautical Laboratory. Here, we describe its first prototype design and report the results of its field tests on the Swiss Morteratsch glacier. Although the IceMole design is currently adapted to terrestrial glaciers and ice shields, it may later be modified for the subsurface in-situ investigation of extraterrestrial ice, e.g., on Mars, Europa, and Enceladus. If life exists on those bodies, it may be present in the ice (as life can also be found in the deep ice of Earth).
Tests with palm tree leaves have just started yet and scan data are in the process to be analyzed. The final goal of future project for palm tree gender and species recognition will be to develop optical scanning technology to be applied to date palm tree leaves for in–situ screening purposes. Depending on the software used and the particular requirements of the users the technology potentially shall be able to identify palm tree diseases, palm tree gender, and species of young date palm trees by scanning leaves.
The demand of replacements for inoperable organs exceeds the amount of available organ transplants. Therefore, tissue engineering developed as a multidisciplinary field of research for autologous in-vitro organs. Such three dimensional tissue constructs request the application of a bioreactor. The UREPLACE bioreactor is used to grow cells on tubular collagen scaffolds OPTIMAIX Sponge 1 with a maximal length of 7 cm, in order to culture in vitro an adequate ureter replacement. With a rotating unit, (urothelial) cells can be placed homogeneously on the inner scaffold surface. Furthermore, a stimulation is combined with this bioreactor resulting in an orientation of muscle cells. These culturing methods request a precise control of several parameters and actuators. A combination of a LabBox and the suitable software LabVision is used to set and conduct parameters like rotation angles, velocities, pressures and other important cell culture values. The bioreactor was tested waterproof successfully. Furthermore, the temperature controlling was adjusted to 37 °C and the CO2 - concentration regulated to 5 %. Additionally, the pH step responses of several substances showed a perfect functioning of the designed flow chamber. All used software was tested and remained stable for several days.
Reasoning with Qualitative Positional Information for Domestic Domains in the Situation Calculus
(2011)
At (ultra)high magnetic fields the artifact sensitivity of ECG recordings increases. This bears the risk of R-wave mis-registration which has been consistently reported for ECG triggered CMR at 7.0T. Realizing the constraints of conventional ECG, acoustic cardiac triggering (ACT) has been proposed. The clinical ACT has not been carefully examined yet. For this reason, this work scrutinizes the suitability, accuracy and reproducibility of ACT for CMR at 7.0T. For this purpose, the trigger reliability and trigger detection variance are examined together with an qualitative and quantitative assessment of image quality of the heart at 7.0T.
One-chip integrated dual amperometric/field-effect sensor for the detection of dissolved hydrogen
(2011)
Lignocellulosic biorefinery: Process integration of hydrolysis and fermentation (SSF process)
(2011)
The aim of the present work is the process integration and the optimization of the enzymatic hydrolysis of wood and the following fermentation of the products to ethanol. The substrate is a fiber fraction obtained by organosolv pre-treatment of beech wood. For the ethanol production, a co-fermentation by two different yeasts (Saccharomyces cerevisiae and Pachysolen tannophilus) was carried out to convert glucose as well as xylose. Two approaches has been followed: 1. A two step process, in which the hydrolysis of the fiber fraction and the fermentation to product are separated from each other. 2. A process, in which the hydrolysis and the fermentation are carried out in one single process step as simultaneous saccharification and fermentation (SSF). Following the first approach, a yield of about 0.15 g ethanol per gram substrate can be reached. Based on the SSF, one process step can be saved, and additionally, the gained yield can be raised up to 0.3 g ethanol per gram substrate.
Bio-feedstocks
(2011)
The magnetic forces of fringe magnetic fields of MR systems on ferromagnetic components can impose a severe patient, occupational health and safety hazard. MRI accidents are listed as number 9 of the top 10 risks in modern medicine. With the advent of ultrahigh field MR systems including passively shielded magnet versions, this risk, commonly known as the missile or projectile effect is even more pronounced. A strategy employing magnetic field sensors which can be attached to ferromagnetic objects that are commonly used in a clinical environment is conceptually appealing for the pursuit of reducing the risk of ferromagnetic projectile accidents.
Absatzkanäle kaum verknüpft
(2011)
Objective
The purpose of this study is to (i) design a small and mobile Magnetic field ALert SEnsor (MALSE), (ii) to carefully evaluate its sensors to their consistency of activation/deactivation and sensitivity to magnetic fields, and (iii) to demonstrate the applicability of MALSE in 1.5 T, 3.0 T and 7.0 T MR fringe field environments.
Methods
MALSE comprises a set of reed sensors, which activate in response to their exposure to a magnetic field. The activation/deactivation of reed sensors was examined by moving them in/out of the fringe field generated by 7TMR.
Results
The consistency with which individual reed sensors would activate at the same field strength was found to be 100% for the setup used. All of the reed switches investigated required a substantial drop in ambient magnetic field strength before they deactivated.
Conclusions
MALSE is a simple concept for alerting MRI staff to a ferromagnetic object being brought into fringe magnetic fields which exceeds MALSEs activation magnetic field. MALSE can easily be attached to ferromagnetic objects within the vicinity of a scanner, thus creating a barrier for hazardous situations induced by ferromagnetic parts which should not enter the vicinity of an MR-system to occur.
Geld-zurück-Garantien erlangen in der Unternehmenspraxis eine immer größere Bedeutung, vor allem weil sie als probates Mittel zur Signalisierung hochwertiger Qualität angesehen werden – eine Annahme, die bislang wissenschaftlich ungeprüft geblieben ist. Vor diesem Hintergrund wird im vorliegenden Beitrag eine umfassende empirische Untersuchung der kaufverhaltensrelevanten Wirkungen dieses Marketinginstrumentes vorgenommen. Die Ergebnisse verdeutlichen zum einen, dass eine Geld-zurück-Garantie nur unter bestimmten Bedingungen als Qualitätssignal wirkt. Dies hängt neben der Art des Produktes (Erfahrungs- vs. Suchgut) insbesondere von der Ausprägung des für die Qualitätsbeurteilung besonders diagnostischen Merkmals Marke sowie von der Produktkenntnis der Konsumenten ab. Zum anderen zeigt sich aber auch, dass eine Geld-zurück-Garantie affektive Konsumentenreaktionen auslöst, die die Kaufabsicht von Konsumenten zusätzlich erhöhen können. Zusammenfassend stellen wir fest, dass eine Geld-zurück-Garantie – entgegen bisheriger Erwartungen aus der Praxis – nicht zwingend ein Qualitätsindikator ist, stattdessen entfaltet sie aber bisher unbeachtete affektive Wirkungen, die insbesondere auf ihre absichernde Funktion von etwaigen Fehlentscheidungen beim Kauf zurückzuführen sind.
Purpose
To design and evaluate a four-channel cardiac transceiver coil array for functional cardiac imaging at 7T.
Materials and Methods
A four-element cardiac transceiver surface coil array was developed with two rectangular loops mounted on an anterior former and two rectangular loops on a posterior former. specific absorption rate (SAR) simulations were performed and a Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-1 calibration method was applied prior to obtain 2D FLASH CINE (mSENSE, R = 2) images from nine healthy volunteers with a spatial resolution of up to 1 × 1 × 2.5 mm3.
Results
Tuning and matching was found to be better than 10 dB for all subjects. The decoupling (S21) was measured to be >18 dB between neighboring loops, >20 dB for opposite loops, and >30 dB for other loop combinations. SAR values were well within the limits provided by the IEC. Imaging provided clinically acceptable signal homogeneity with an excellent blood-myocardium contrast applying the Burn:x-wiley:10531807:media:JMRI22451:tex2gif-stack-2 calibration approach.
Conclusion
A four-channel cardiac transceiver coil array for 7T was built, allowing for cardiac imaging with clinically acceptable signal homogeneity and an excellent blood-myocardium contrast. Minor anatomic structures, such as pericardium, mitral, and tricuspid valves and their apparatus, as well as trabeculae, were accurately delineated.