Finite element modelling of stapled colorectal end-to-end anastomosis : Advantages of variable height stapler design

  • The impact of surgical staplers on tissues has been studied mostly in an empirical manner. In this paper, finite element method was used to clarify the mechanics of tissue stapling and associated phenomena. Various stapling modalities and several designs of circular staplers were investigated to evaluate the impact of the device on tissues and mechanical performance of the end-to-end colorectal anastomosis. Numerical simulations demonstrated that a single row of staples is not adequate to resist leakage due to non-linear buckling and opening of the tissue layers between two adjacent staples. Compared to the single staple row configuration, significant increase in stress experienced by the tissue at the inner staple rows was observed in two and three rows designs. On the other hand, adding second and/or third staple row had no effect on strain in the tissue inside the staples. Variable height design with higher staples in outer rows significantly reduced the stresses and strains in outer rows when compared to the same configuration with flat cartridge.

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar
Metadaten
Verfasserangaben:V. Novacek, Thanh Ngoc TranORCiD, U. Klinge, R. H. Tolba, Manfred StaatORCiD, D. G. Bronson, A. M. Miesse, J. Whiffen, F. Turquier
DOI:https://doi.org/10.1016/j.jbiomech.2012.07.021
ISSN:1873-2380
Titel des übergeordneten Werkes (Englisch):Journal of Biomechanics
Verlag:Elsevier
Verlagsort:Amsterdam
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Erscheinungsjahr:2012
Datum der Publikation (Server):18.12.2012
Freies Schlagwort / Tag:Anastomotic leakage; End-to-end colorectal anastomosis; Finite element modelling; Surgical staplers; Variable height stapler design
Jahrgang:45
Ausgabe / Heft:115
Erste Seite:2693
Letzte Seite:2697