• Treffer 19 von 0
Zurück zur Trefferliste

FEM-based modeling of a calorimetric gas sensor for hydrogen peroxide monitoring

  • A physically coupled finite element method (FEM) model is developed to study the response behavior of a calorimetric gas sensor. The modeled sensor serves as a monitoring device of the concentration of gaseous hydrogen peroxide (H2 O2) in a high temperature mixture stream in aseptic sterilization processes. The principle of operation of a calorimetric H2 O2 sensor is analyzed and the results of the numerical model have been validated by using previously published sensor experiments. The deviation in the results between the FEM model and experimental data are presented and discussed.

Metadaten exportieren

Weitere Dienste

Teilen auf X Suche bei Google Scholar
Metadaten
Verfasserangaben:Zaid B. Jildeh, Patrick Kirchner, Jan OberländerORCiD, Alexander Kremers, Torsten WagnerORCiD, Patrick H. Wagner, Michael Josef SchöningORCiD
DOI:https://doi.org/10.1002/pssa.201600912
ISSN:1862-6319
Titel des übergeordneten Werkes (Englisch):physica status solidi a : applications and materials sciences
Verlag:Wiley-VCH
Verlagsort:Weinheim
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Erscheinungsjahr:2017
Datum der Publikation (Server):21.08.2017
Ausgabe / Heft:Early View
Link:https://doi.org/10.1002/pssa.201600912
Zugriffsart:campus
Fachbereiche und Einrichtungen:FH Aachen / Fachbereich Medizintechnik und Technomathematik
FH Aachen / INB - Institut für Nano- und Biotechnologien
collections:Verlag / Wiley-VCH