Detection of Adrenaline in Blood Plasma as Biomarker for Adrenal Venous Sampling

  • An amperometric bi-enzyme biosensor based on substrate recycling principle for the amplification of the sensor signal has been developed for the detection of adrenaline in blood. Adrenaline can be used as biomarker verifying successful adrenal venous sampling procedure. The adrenaline biosensor has been realized via modification of a galvanic oxygen sensor with a bi-enzyme membrane combining a genetically modified laccase and a pyrroloquinoline quinone-dependent glucose dehydrogenase. The measurement conditions such as pH value and temperature were optimized to enhance the sensor performance. A high sensitivity and a low detection limit of about 0.5–1 nM adrenaline have been achieved in phosphate buffer at pH 7.4, relevant for measurements in blood samples. The sensitivity of the biosensor to other catecholamines such as noradrenaline, dopamine and dobutamine has been studied. Finally, the sensor has been successfully applied for the detection of adrenaline in human blood plasma.

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar
Metadaten
Verfasserangaben:Denise MolinnusORCiD, Gabriel Hardt, Petra SiegertORCiD, Holger S. Willenberg, Arshak PoghossianORCiD, Michael Keusgen, Michael Josef SchöningORCiD
DOI:https://doi.org/10.1002/elan.201800026
ISSN:1521-4109
Titel des übergeordneten Werkes (Englisch):Electroanalysis
Verlag:Wiley-VCH
Verlagsort:Weinheim
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Erscheinungsjahr:2018
Datum der Publikation (Server):19.02.2018
Jahrgang:30
Ausgabe / Heft:5
Erste Seite:937
Letzte Seite:942
Link:http://doi.org/10.1002/elan.201800026
Zugriffsart:campus
Fachbereiche und Einrichtungen:FH Aachen / Fachbereich Medizintechnik und Technomathematik
FH Aachen / INB - Institut für Nano- und Biotechnologien
collections:Verlag / Wiley-VCH