Refine
Year of publication
Document Type
- Article (1299)
- Conference Proceeding (131)
- Book (43)
- Part of a Book (40)
- Doctoral Thesis (18)
- Other (5)
- Patent (4)
- Preprint (3)
- Habilitation (1)
- Talk (1)
Language
- English (1545) (remove)
Has Fulltext
- no (1545) (remove)
Keywords
- LAPS (4)
- Natural language processing (4)
- CellDrum (3)
- Field-effect sensor (3)
- Light-addressable potentiometric sensor (3)
- Paired sample (3)
- hydrogen peroxide (3)
- Bacillus atrophaeus (2)
- Biocomposites (2)
- Clustering (2)
- Empirical process (2)
- Force (2)
- Goodness-of-fit test (2)
- Incomplete data (2)
- Independence test (2)
- Information extraction (2)
- Iterative learning control (2)
- Limit analysis (2)
- Machine learning (2)
- Natural fibres (2)
Institute
- Fachbereich Medizintechnik und Technomathematik (1545) (remove)
Cardiopulmonary bypass (CPB) is a standard technique for cardiac surgery, but comes with the risk of severe neurological complications (e.g. stroke) caused by embolisms and/or reduced cerebral perfusion. We report on an aortic cannula prototype design (optiCAN) with helical outflow and jet-splitting dispersion tip that could reduce the risk of embolic events and restores cerebral perfusion to 97.5% of physiological flow during CPB in vivo, whereas a commercial curved-tip cannula yields 74.6%. In further in vitro comparison, pressure loss and hemolysis parameters of optiCAN remain unaffected. Results are reproducibly confirmed in silico for an exemplary human aortic anatomy via computational fluid dynamics (CFD) simulations. Based on CFD simulations, we firstly show that optiCAN design improves aortic root washout, which reduces the risk of thromboembolism. Secondly, we identify regions of the aortic intima with increased risk of plaque release by correlating areas of enhanced plaque growth and high wall shear stresses (WSS). From this we propose another easy-to-manufacture cannula design (opti2CAN) that decreases areas burdened by high WSS, while preserving physiological cerebral flow and favorable hemodynamics. With this novel cannula design, we propose a cannulation option to reduce neurological complications and the prevalence of stroke in high-risk patients after CPB.
Optoelectronic Properties of Nanostructured Ensembles Controlled by Biomolecular Logic Systems
(2008)
Purpose
The aim of this study was to compare several osteosynthesis techniques (intramedullary headless compression screws, T-plates, and Kirschner wires) for distal epiphyseal fractures of proximal phalanges in a human cadaveric model.
Methods
A total of 90 proximal phalanges from 30 specimens (index, ring, and middle fingers) were used for this study. After stripping off all soft tissue, a transverse distal epiphyseal fracture was simulated at the proximal phalanx. The 30 specimens were randomly assigned to 1 fixation technique (30 per technique), either a 3.0-mm intramedullary headless compression screw, locking plate fixation with a 2.0-mm T-plate, or 2 oblique 1.0-mm Kirschner wires. Displacement analysis (bending, distraction, and torsion) was performed using optical tracking of an applied random speckle pattern after osteosynthesis. Biomechanical testing was performed with increasing cyclic loading and with cyclic load to failure using a biaxial torsion-tension testing machine.
Results
Cannulated intramedullary compression screws showed significantly less displacement at the fracture site in torsional testing. Furthermore, screws were significantly more stable in bending testing. Kirschner wires were significantly less stable than plating or screw fixation in any cyclic load to failure test setup.
Conclusions
Intramedullary compression screws are a highly stable alternative in the treatment of transverse distal epiphyseal phalangeal fractures. Kirschner wires seem to be inferior regarding displacement properties and primary stability.
Clinical relevance
Fracture fixation of phalangeal fractures using plate osteosynthesis may have the advantage of a very rigid reduction, but disadvantages such as stiffness owing to the more invasive surgical approach and soft tissue irritation should be taken into account. Headless compression screws represent a minimally invasive choice for fixation with good biomechanical properties.
Our world is well ordered in measurement and number : or why natural constants are as they are
(2013)
All the important natural constants can be logically explained with and derived from the first four ordinal numbers, 1, 2, 3 and 4, its addition to ten and finally the standard values for obviously maximal feasibility Ω and the optimum in our world, the Golden Section (GS), i.e. the number sequences 273 and 618. They both are the first three numbers of irrational results by an arithmetical transformation of simple geometrical relationships by creating multiplicity out of singularity. Both of them show that the infinite is inherent in finiteness and explain in a simple way the smallest deviations and fluctuations between the physical AS-IS state and the obvious spiritual ideal behind: Wherever we look in this world, and especially in important key-positions, we regularly find these sequences. All of the above mentioned numbers so seem to be key players in our world, what can be demonstrated by the derivation of natural constants.
The vaginal prolapse after hysterectomy (removal of the uterus) is often associated with the prolapse of the vaginal vault, rectum, bladder, urethra or small bowel. Minimally
invasive surgery such as laparoscopic sacrocolpopexy and pectopexy are widely performed for the treatment of the vaginal prolapse with weakly supported vaginal vault after hysterectomy using prosthetic mesh implants to support (or strengthen) lax apical ligaments. Implants of different shape, size and polymers are selected depending on the patient’s anatomy and the surgeon’s preference. In this computational study on pectopexy, DynaMesh®-PRP soft, GYNECARE GYNEMESH® PS Nonabsorbable PROLENE® soft and Ultrapro® are tested in a 3D finite element model of the female pelvic floor. The mesh model is implanted into the extraperitoneal space and sutured to the vaginal stump with a bilateral fixation to the iliopectineal ligament at both sides. Numerical simulations are conducted at rest, after surgery and during Valsalva maneuver with weakened tissues modeled by reduced tissue stiffness. Tissues and prosthetic meshes are modeled as incompressible, isotropic hyperelastic materials. The positions of the organs are calculated with respect to the pubococcygeal line (PCL) for female pelvic floor at rest, after repair and during Valsalva maneuver using the three meshes.
Penicillin detection by means of field-effect based sensors: EnFET, capacitive EIS sensor or LAPS?
(2001)