The ClearPET™ project: Development of a 2nd generation high-performance small animal PET scanner
(2005)
Second generation high-performance PET scanners, called ClearPET™1, have been developed by working groups of the Crystal Clear Collaboration (CCC). High sensitivity and high spatial resolution for the ClearPET camera is achieved by using a phoswich arrangement combining two different types of lutetium-based scintillator materials: LSO from CTI and LuYAP:Ce from the CCC (ISTC project). In a first ClearPET prototype, phoswich arrangements of 8×8 crystals of 2×2×10 mm3 are coupled to multi-channel photomultiplier tubes (Hamamatsu R7600). A unit of four PMTs arranged in-line represents one of 20 sectors of the ring design. The opening diameter of the ring is 120 mm, the axial detector length is 110 mm.The PMT pulses are digitized by free-running ADCs and digital data processing determines the gamma energy, the phoswich layer and even the exact pulse starting time, which is subsequently used for coincidence detection. The gantry allows rotation of the detector modules around the field of view.
Preliminary data shows a correct identification of the crystal layer about (98±1)%. Typically the energy resolution is (23.3±0.5)% for the luyap layer and (15.4±0.4)% for the lso layer. early studies showed the timing resolution of 2 ns FWHM and 4.8 ns FWTM. the intrinsic spatial resolution ranges from 1.37 mm to 1.61 mm full-width of half-maximum (FWHM) with a mean of 1.48 mm FWHM. further improvements in image and energy resolution are expected when the system geometry is fully modeled.
The Crystal Clear Collaboration has developed a modular system for a small animal PET scanner (ClearPET). The modularity allows the assembly of scanners of different sizes and characteristics in order to satisfy the specific needs of the individual member institutions. The system performs depth of interaction detection by using a phoswich arrangement combining LSO and LuYAP scintillators which are coupled to Multichannel Photomultipliers (PMTs). For each PMT a free running 40 MHz ADC digitizes the signal and the complete scintillation pulse is sampled by an FPGA and sent with 20 MB/s to a PC for preprocessing. The pulse provides information about the gamma energy and the scintillator material which identifies the interaction layer. Furthermore, the exact pulse starting time is obtained from the sampled data. This is important as no hardware coincidence detection is implemented. All single events are recorded and coincidences are identified by software. The system in Jülich (ClearPET Neuro) is equipped with 10240 crystals on 80 PMTs. The paper will present an overview of the data acquisition system.
The small animal PET scanners developed by the Crystal Clear Collaboration (ClearPETtrade) detect coincidences by analyzing timemarks which are attached to each event. The scanners are able to save complete single list mode data which allows analysis and modification of the timemarks after data acquisition. The timemarks are obtained from the digitally sampled detector pulses by calculating the baseline crossing of the rising edge of the pulse which is approximated as a straight line. But the limited sampling frequency causes a systematic error in the determination of the timemark. This error depends on the phase of the sampling clock at the time of the event. A statistical method that corrects these errors will be presented